4. Rotate the clamping ring clockwise until the slot aligns with the tab on the screw gear. Remove the back-end lock by sliding
it off the screw gear (Figure 19).
5. Spread the screw gear halves apart to gain access to the delivery system inner lumens (Figure 20).
6. Using hemostats, grip the guidewire tube adjacent to the winged tip-release fitting. Using another hemostats, grip the
broken outer lumen adjacent to the end of the tube (Figure 21).
Figure 18. Separate the handle halves
Figure 19. Removing the back-end lock
Figure 20. Spreading the screw gear halves
Figure 21. Gripping the guidewire tube (right) and tip capture tube (left)
Instructions for Use