Guidance and manufacturer's declaration – electromagnetic immunity
The KARL STORZ NBU module 408100 20 is intended for use in the electromagnetic environment specified below.
The customer or user of the KARL STORZ NBU module 408100 20 should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
± 2 kV for power supply
Electrical fast
lines
transient/burst
± 1 kV for input/output
IEC 61000-4-4
lines
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
<5 % U
(>95 % dip in U
for 0.5 cycle
40 % UT
(60 % dip in U
Voltage dips, short interrup-
for 5 cycles
tions and voltage variations on
power supply input lines
70 % U
IEC 61000-4-11
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Magnetic field at the mains
frequency (50/60 Hz)
3 A/m
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Appendix
Information
Table 2
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5 % U
*
T
T
)
(>95 % dip in U
)
T
T
for 0.5 cycle
Complies
40 % U
T
)
(60 % dip in U
)
T
T
for 5 cycles
Complies
T
70 % U
)
T
T
(30 % dip in U
)
T
for 25 cycles
Complies
T
<5 % U
)
T
T
(>95 % dip in U
)
T
for 5 sec
Complies
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical commercial
or hospital environment. If image distortion occurs it may be
necessary to further filter the power quality by using a AC line
filter.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
If the user of the equipment or system requires continued
operation during power mains interruptions, it is recommended
that the equipment or system be powered from an uninterruptible
power supply or a battery.
The magnetic fields at the mains frequency should be at the
values typical of business or hospital environments.