[Start Calibration]:
1.
[Call the Main menu]:
2.
[Select Configuration]:
3.
Use the scrollbar on the right of the screen to scroll to [Calibration].
4.
[Select Calibration]
ð Screen changes to [Calibration].
5.
'Press Start' .
ð Screen changes to 'Calibration step 1' .
6.
Calibration step 1 runs automatically.
The calibration runs automatically in this step. The pump completes a
given number of pump-specific strokes and counts them down to 0. At
the same time, the oval gear meter (OGM) pulses are also counted. The
pump stops after the strokes elapse.
7.
Screen 'Calibration step 1' is launched.
Use the slider to enter the quantity in litres (before the decimal point).
8.
Select the quantity in millilitres after the decimal point.
9.
Use the slider to enter the quantity in millilitres (after the decimal point).
10. [Press the Next] button
ð Screen changes to the 'Calibration step 2' .
11.
'Calibration step 2' runs automatically.
The calibration sequence runs again automatically in this step.
12. Screen 'Calibration step 2' is launched.
Use the slider to enter the quantity in litres (before the decimal point).
13. Select the quantity in millilitres after the decimal point.
14. Use the slider to enter the quantity in millilitres (after the decimal point).
15. [Press the Next] button
ð The setting is saved, and the screen switches back to the operation screen.
Ä Chapter 8.5 'Main menu' on page 63
Ä Chapter 8.8 'Configuration' on page 89
.
.
125
Control/software
MAN046939 Rev. 6-10.2021