LOGic:GROup<m>:TECHnology <ThresholdType>
Sets the threshold value for the selected channel group, or for all logic channels.
Suffix:
<m>
Parameter:
<ThresholdType>
LOGic:GROup<m>:USER <ThresholdValue>
Sets the threshold value if
Suffix:
<m>
Benutzerhandbuch 1326.1578.03 ─ 19
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if
TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
TTL
1.4 V
ECL
-1.3 V
CMOS
2.5 V
GND
0 V: for CAN channels, requires the option R&S RTH-K3.
CAN
2 V: for CAN channels, requires the option R&S RTH-K3.
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V: for LIN channels, requires the option
R&S RTH-K3.
12 V Supply
12 V: for LIN channels, requires the option R&S RTH-K3.
18 V Supply
18 V: for LIN channels, requires the option R&S RTH-K3.
USER
Set the value with LOGic:GROup<m>:USER.
*RST:
TTL
LOGic:GROup<m>:TECHnology
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if
Fernsteuerbefehle
Logikanalysator (R&S RTH-B1 MSO)
LOGic:THCoupling
is set to USER.
LOGic:THCoupling
is OFF.
is OFF.