R&S
Scope Rider RTH
Parameter:
<NoOfAvgs>
ACQuire:ARESet:IMMediate
Restarts the envelope and average calculation.
Verwendung:
ACQuire:POINts:PRESelect <RecLength>
Sets the record length.
Parameter:
<RecLength>
The actual record length can differ from these limits, see
- C4 (Act. Record Len. C1 - C4)"
ACQuire:WAVeformupd <WaveformUpdate>
The command is relevant if the time scale is ≥50 ms/div. At these slow timebases, you
can select how the acquired samples are displayed.
Parameter:
<WaveformUpdate> INTermediate | FULL
Benutzerhandbuch 1326.1578.03 ─ 19
Bereich:
2 bis 8192
Inkrement: 2
N
(N = 1 .. 13)
*RST:
2
Ereignis
MAX | MIDDle | MIN
MAX
Sets the maximum record length.
MIDDle
Limits the record length to 12.5 ksample.
MIN
Limits the record length to 1.25 ksample.
*RST:
MAX
auf Seite 55.
INTermediate
The acquired samples are displayed before the acquisition is
completed. In this mode, the record length is limited to
125 ksample.
FULL
The acquired samples are displayed when the complete acquisi-
tion has been recorded. Depending on the selected time scale, it
takes some time until the waveform is visible. This mode does
not limit the record length and is always used for time scales
<50 ms/div.
*RST:
INTermediate
Fernsteuerbefehle
Einrichten von Messkurven
„Akt. Aufzeichnungslänge C1
330