15.8.2.4
AC+DC Strom- und Spannungsmessungen
METer<m>:SENSe:CURRent:AC:COUPling
METer<m>:SENSe:VOLTage:AC:COUPling
METer<m>:SENSe:CURRent:AC:COUPling <Coupling>
METer<m>:SENSe:VOLTage:AC:COUPling <Coupling>
Enables AC+DC measurements.
Suffix:
<m>
Parameter:
<CoupVoltageMeas> DCLimit | ACLimit
Benutzerhandbuch 1326.1578.03 ─ 19
.
1..4
Selects the voltmeter.
<numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1000 V.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Bereich:
1 bis 1000
*RST:
1
Std.-einheit: V
..................................................................... 421
...................................................................... 421
.
1..4
R&S RTH1002: always 1, omit the suffix.
R&S RTH1004: Selects the voltmeter.
DCLimit
Enables AC+DC measurement.
ACLimit
Enables AC measurement.
*RST:
DCLimit
Fernsteuerbefehle
Voltmeter (R&S RTH1004)