®
R&S
Scope Rider RTH
Parameter:
<Scaling>
SPECtrum:FREQuency:MAGNitude:SCALe <MagnitudeUnit>
Queries or defines the scale unit for the amplitude range (y-axis) in the spectrum dis-
play.
In FFT mode, this command is query-only.
Parameter:
<MagnitudeUnit>
SPECtrum:FREQuency:SAMPle?
Queries the sample rate with which FFT analysis is performed. This command is only
available in FFT mode.
Rückgabewerte:
<FFTSmpFreq>
Verwendung:
SPECtrum:FREQuency:SPAN:MODE <Mode>
Defines how many values are analyzed by a single FFT, and thus the frequency resolu-
tion.
Parameter:
<Mode>
SPECtrum:FREQuency:SPAN[:VALue] <HorizontalSpan>
Queries or defines the used span, which is: how many values are analyzed by a single
FFT.
In FFT mode, this command is query-only.
Benutzerhandbuch 1326.1578.03 ─ 19
LINear | LOGarithmic
*RST:
LINear
DBM | DBV | DBA
*RST:
DBM
Bereich:
1 bis 10E+9
Inkrement: 10
*RST:
1
Std.-einheit: Hz
Nur Abfrage
NARRow | MAX
NARRow
8k values analyzed per FFT; time scale ≥ 100 ms/div
MAX
64k values analyzed per FFT; time scale ≥ 1 us/div
*RST:
MAX
Fernsteuerbefehle
Spektrumanalyse
376