Herunterladen Diese Seite drucken

Rohde & Schwarz Scope Rider RTH Benutzerhandbuch Seite 333

Handheld-oszilloskop
Vorschau ausblenden Andere Handbücher für Scope Rider RTH:
®
R&S
Scope Rider RTH
TRIGger:TYPE <Type>
Selects the trigger type, the event type that defines the trigger point.
Parameter:
<Type>
TRIGger:LEVel<m>:VALue <Level>
Sets the trigger level voltage. The command is relevant for all trigger types that require
one trigger level.
Suffix:
<m>
Parameter:
<Level>
TRIGger:HOLDoff:MODE <Mode>
Selects the method to define the holdoff.
Parameter:
<Mode>
Benutzerhandbuch 1326.1578.03 ─ 19
EDGE | GLITch | WIDTh | TV | PATTern | STATe | RUNT |
SLEWrate | DATatoclock | SERPattern | TIMeout | INTerval |
WINDow | PROTocol
EDGE | GLITch | WIDTh | TV
Standard trigger types
PATTern | STATe | RUNT | SLEWrate | DATatoclock | SERPat-
tern | TIMeout | INTerval | WINDow
Require option R&S RTH-K19.
PROTocol
Requires option R&S RTH-K1 and/or R&S RTH-K2
*RST:
EDGE
.
1..21
Indicates the trigger source:
1..4: analog channels 1 to 4
8..15: digital channels D0 to D7
All others: not available
Bereich:
-10 bis 10
Inkrement: 1E-3
*RST:
0
Std.-einheit: V
OFF | TIME | RANDom | EVENts
OFF
No holdoff
TIME
Defines the holdoff as a time period. The next trigger occurs only
after a time has passed, which is defined with
HOLDoff:TIME.
Fernsteuerbefehle
Einrichten von Messkurven
TRIGger:
333

Kapitel

loading