Beispiel:
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:DEViation <Deviation>
Sets the distance between the lower range and the mean value and between the upper
range and the mean value.
The command disables the vertical auto scaling.
See also: LOGGer:MEASurement<m>:VERTical:MEAN.
Suffix:
<m>
Parameter:
<Deviation>
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:MEAN <Mean>, [<Deviation>]
Sets the mean value of the scaling range. In addition, you can set the deviation. The
query returns the mean value.
The command disables the vertical auto scaling.
Use this command if you want to define the range based on the mean value and the
standard deviation, see
MEASurement<m>:RESult:STDDev?
Suffix:
<m>
Parameter:
<Mean>
Einstellparameter:
<Deviation>
Beispiel:
Benutzerhandbuch 1326.1578.03 ─ 19
Set the lower value to -1 and the upper value to +1 using one
command:
LOGGer:MEASurement1:VERTical:LOWer -1, +1
//or
LOGGer:MEASurement1:VERTical:UPPer +1, -1
.
1..4
Logger channel
Absolute difference between mean value and upper/lower range
Bereich:
1E-15 bis 100E+24
Inkrement: 1
*RST:
0.707
LOGGer:MEASurement<m>:RESult:MEAN?
auf Seite 446.
.
1..4
Logger channel
Middle value between the upper and lower scaling value.
Absolute difference between mean value and upper/lower sca-
ling value
LOGGer:MEASurement1:VERTical:MEAN 10, 1
Sets the mean value to 10, and the deviation to 1. Thus, the
upper range is 10 + 1 = 11, and the lower range is 10 - 1 = 9.