R&S
Scope Rider RTH
Parameter:
<HorizontalSpan>
SPECtrum:FREQuency:STARt <HorizontalStart>
Queries or defines the start frequency of the used span. The value depends on the
SPECtrum:FREQuency:SPAN:MODE
In FFT mode, this command is query-only.
Parameter:
<HorizontalStart>
SPECtrum:FREQuency:STOP <HorizontalStop>
Queries or defines the stop frequency of the used span. The value depends on the
SPECtrum:FREQuency:SPAN:MODE
In FFT mode, this command is query-only.
Parameter:
<HorizontalStop>
SPECtrum:FREQuency:WINDow:FACTor?
Queries the window factor used for FFT analysis.
Rückgabewerte:
<WindowFactor>
Verwendung:
SPECtrum:FREQuency:WINDow:TYPE <WindowType>
Determines the window function used for FFT analysis. See
Parameter:
<WindowType>
Benutzerhandbuch 1326.1578.03 ─ 19
8E+6 | 64E+6
Bereich:
1000 bis 500E+6
Inkrement: 10
*RST:
500E+6
Std.-einheit: Hz
and
CHANnel<m>:BANDwidth
Bereich:
0 bis 499.999E+6
Inkrement: 10
*RST:
0
Std.-einheit: Hz
and
CHANnel<m>:BANDwidth
Bereich:
1000 bis 500E+6
Inkrement: 10
*RST:
500E+6
Std.-einheit: Hz
Bereich:
0.89 bis 3.84
Inkrement: 0.1
*RST:
1.44
Nur Abfrage
RECTangular | FLATtop | HAMMing | HANN | BLACkman
*RST:
HANN
Fernsteuerbefehle
Spektrumanalyse
parameters.
parameters.
Tabelle 6-1
for details.
377