Herunterladen Diese Seite drucken

Rohde & Schwarz Scope Rider RTH Benutzerhandbuch Seite 433

Handheld-oszilloskop
Vorschau ausblenden Andere Handbücher für Scope Rider RTH:
®
R&S
Scope Rider RTH
Suffix:
<m>
Verwendung:
COUNter<m>:MEASure:FREQuency? <Range>
This command sets all measurement parameters and trigger parameters to default
values for frequency measurements, then immediately triggers a counter measurement
and returns the result.
This command corresponds to the commands
FREQuency
Suffix:
<m>
Parameter:
<Range>
Rückgabewerte:
<Counter result>
Verwendung:
COUNter<m>:READ?
Starts a new measurement, returns the current measurement result and stays in run
mode.
Suffix:
<m>
Rückgabewerte:
<Counter result>
Verwendung:
Benutzerhandbuch 1326.1578.03 ─ 19
.
1..2
Ereignis
+
COUNter<m>:FETCh?
.
1..2
<numeric value> | MIN | MAX | DEF
Defines the measurement range of the input signal for the coun-
ter. If probes are connected, the MIN | MAX | DEF values are
adapted accordingly.
<numeric value>
Power range in V. If necessary, the next higher range is selec-
ted.
MIN
100 mV
MAX
300 V
DEF
100 mV
Measured frequency
Nur Abfrage
.
1..2
Measured frequency
Nur Abfrage
Fernsteuerbefehle
Frequenzzähler (R&S RTH-K33)
COUNter<m>:CONFigure: