R&S
Scope Rider RTH
ACQuire:POINts:PRESelect
ACQuire:WAVeformupd
ACQuire:POINts:ARATe?
ACQuire:POINts[:VALue]?
ACQuire:RESolution?
RUN
Starts the continuous acquisition.
Verwendung:
STOP
Stops the running acquisition.
Verwendung:
ACQuire:MODE <AcquisitionMode>
Defines how the waveform is built from the captured samples.
Parameter:
<AcquisitionMode>
ACQuire:AVERage:COUNt <NoOfAvgs>
Sets the number of waveforms used to calculate the average waveform.
Benutzerhandbuch 1326.1578.03 ─ 19
............................................................................................ 330
.................................................................................................. 330
................................................................................................ 331
...............................................................................................331
..................................................................................................... 331
Ereignis
Ereignis
SAMPle | PDETect | HRESolution | AVERage | ENVelope
SAMPle
One of n samples in a sample interval is recorded as waveform
point, the other samples are discarded.
PDETect
The minimum and the maximum of n samples are recorded as
waveform points, the other samples are discarded.
HRESolution
The average of n captured sample points is recorded as one
waveform point.
AVERage
The average is calculated from the data of the current acquisi-
tion and a number of acquisitions before. The number of acquisi-
tions for average calculation is defined with
AVERage:COUNt.
ENVelope
The minimum and maximum values in a sample interval over a
number of acquisitions are saved. The most extreme values of
all acquisitions build the envelope.
*RST:
SAMPle
Fernsteuerbefehle