15.10.3 Cursor im Logger-Modus

15.10.3.1
Cursor-Einstellungen
LOGGer:CURSor[:STATe]
LOGGer:CURSor:TYPE
LOGGer:CURSor:SCPLing
LOGGer:CURSor:COUPling
LOGGer:CURSor:SCReen
LOGGer:CURSor[:STATe] <Enabled>
Enables or disables the logger cursor measurement.
Parameter:
<Enabled>
LOGGer:CURSor:TYPE <Value>
If data logging is running for more than 2 Tage und 7 Stunden, the logger compresses
4 logging values into a minimum, average and maximum value. The command defines
which of these values are measured at the cursor position.
For logging periods shorter than 2 Tage und 7 Stunden, the command is not relevant.
Parameter:
<Value>
LOGGer:CURSor:SCPLing <TrackScaling>
If ON, the position of the cursor lines is adjusted when the horizontal scale is changed.
If OFF, the cursor lines remain on their position on the display when the scaling is
changed.
Parameter:
<TrackScaling>
LOGGer:CURSor:COUPling <Coupling>
Couples the cursor lines so that the distance between the two lines remains the same
when one cursor is moved.
Benutzerhandbuch 1326.1578.03 ─ 19
............................................................................................... 441
..................................................................................................441
............................................................................................. 441
............................................................................................441
..............................................................................................442
ON | OFF
*RST:
OFF
MINimum | AVERage | MAXimum
*RST:
AVERage
ON | OFF
*RST:
OFF
Fernsteuerbefehle
Daten-Logger
441