Herunterladen Diese Seite drucken

Bus:can:data:threshold; Bus:setreflevels; Bus:can:samplepoint; Bus:can:fdata:abitrate - Rohde & Schwarz Scope Rider Rth Benutzerhandbuch

Handheld-oszilloskop
Vorschau ausblenden Andere Handbücher für Scope Rider RTH:
®
R&S
Scope Rider RTH
BUS:CAN:DATA:THReshold <ThresholdValue>
Sets the threshold value for digitization of signals manually. If the signal value is higher
than the threshold, the signal state is high. Otherwise, the signal state is considered
low.
This value is only considered for
Parameter:
<ThresholdValue>

BUS:SETReflevels

Sets the appropriate threshold or threshold for the selected serial protocol.
Verwendung:
BUS:CAN:SAMPlepoint <SamplePoint>
The sample point divides the nominal bit period into two distinct time segments. The
length of the time segments is defined in time quanta according to network and node
conditions during CAN development.
Parameter:
<SamplePoint>
BUS:CAN:FDATa:ABITrate <ArbCustomBitrate>
Sets the bit rate of the arbitration phase of the CAN FD bus. Die Einstellung ist in der
CAN-FD-Option R&S RTH-K9 verfügbar.
Parameter:
<ArbCustomBitrate> Bereich:
BUS:CAN:FDATa:ASAMplepoint <ArbSamplePoint>
The sample point divides the arbitration phase into two distinct time segments. The
length of the time segments is defined in time quanta according to network and node
conditions during CAN FD development. Die Einstellung ist in der CAN-FD-Option
R&S RTH-K9 verfügbar.
Benutzerhandbuch 1326.1578.03 ─ 19
BUS:CAN:TECHnology
Bereich:
-400 bis 400
Inkrement: 1E-3
*RST:
1.4
Std.-einheit: V
Ereignis
Bereich:
10 bis 95
Inkrement: 1
*RST:
50
Std.-einheit: %
10000 bis 1000000
Inkrement: 1
*RST:
50000
Std.-einheit: bps
Fernsteuerbefehle
Protokollanalyse
USER
475

Kapitel

loading