5. Place the lid on top of the base. The edge of the base will fit in the lid
channel creating a knife edge fit.
E
6. Secure the closure by latching the lid to the base. The top of the latch fits
n
over the ridge in the lid. Push the bottom section of the latch over the lock
holder. You will feel a solid click.
g
l
7. Place the appropriate metal ID tags in the label holders located on either
i
side of the Container latches. The label holder on the right can
accommodate a load card available from Case Medical, Inc®. Only clear ID
s
tags can be used in H2O2 sterilization.
h
8. Thread the guide on the SteriTite® tamper- evident seal through the lock
holder and secure. Repeat on both latches. Blue and red tamper evident seals are available for steam and
gas. White tamper evident seals are recommended for H2O2/ STERRAD
Sterilization.
Caution: Use of any non-approved tamper evident seal could damage the
locking clip.
9. An external indicator or load card should be attached to the Container at
this time. Case Medical provides external indicators for steam and EO, as
well as H2O2 and gas plasma sterilization.
10. SteriTite Containers are designed to be dry after sterilization. Therefore,
the use of absorbent liners is not recommended for use with the SteriTite®
container.
SteriTite® Sterilization
1. Place the SteriTite® container flat on the shelf of the sterilizer cart. If
needed, up to three (3) Containers may be stacked and processed in an
autoclave.
2. If sterilized in a mix load, place Containers below wrapped or linen items.
3. Consult the recommendations of your sterilizer manufacturer to determine
the correct parameters regarding temperature, weight load, dry time,
instrument processing and pre-and post-conditioning cycles.
Note: To minimize the potential of condensate formation, crack the door of
the autoclave for 10 to 15 mins.
4. Following the steam sterilization process, the cart should be removed
from the autoclave and placed in cool down.
SteriTite® Labeling for Steam Sterilization
PRE-VACUUM STEAM TERMINAL STERILIZATION PARAMETERS FOR USE:
Use vented or solid base Container for pre-vacuum steam. Apply paper or polypropylene disposable filter
for each use. Use MediTray® inserts in the Container to secure instrumentation. Recommended for
sterilization of medical devices including, blades and metal lumens of 2mm minimum diameter up to
435mm in length and porous lumens 3mm minimum diameter up to 400mm in length.
Recommended exposure time: 4 minutes at 270°F.
Recommended dry times:
A minimum of 5 minutes for perforated bottom units
A minimum of 8 minutes for solid bottom units
20 minutes may be required for items stored for later use
Copyright 2020 Case Medical, Inc.®
7
IFU-CASETRANS Rev. 004A