After connecting via the browser, you are initially at the Operator user level
(Symbol
). Click on the icon at the top right of the window to select a different
user level: Maintenance (Symbol
After making the ClipX device settings in your browser, save them all to
your PC by choosing Device Storage from the menu.
7.4.1

Assistant to measure sensors

The Assistant menu option will help you to measure your sensor if you do not
have a calibration protocol or you want to measure the sensor when installed.
Before measuring, set the sensor type in the Amplifier menu to enable the
ClipX to perform measurements. Also specify a suitable low-pass filter, e.g.
10 Hz for static measuring, so as to obtain the most stable possible measure-
ment values. For dynamic measuring, you must select the filter matching your
signal frequencies and process.
The assistant provides two measurement variants:
1. Mode: Static
Two points are measured: the value with the sensor under no load and the
value with the sensor under load (force/pressure/torque, etc.). The mea-
surement itself is performed over a period of about 6 seconds with the filter
set in Amplifier. The calculated mean value and the (sample) standard
deviation are displayed both during and after the measurement process.
2. Mode: Dynamic
This mode is suitable when you are not able to apply a constant load
(force/pressure/torque, etc.). The peak values are calculated over your
measuring time and displayed.
You can either enter the physical value of the load manually or import it over the
ClipX bus or via one of the interfaces.
ClipX
) or Administrator (Symbol
A04643_04_X00_03 HBM: public
Starting up the ClipX
).
85