Herunterladen Inhalt Inhalt Diese Seite drucken

HBM ClipX Bedienungsanleitung Seite 48

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Electrical connections, LEDs
6.6.3
Potentiometric transducer
See also
"Available connections and LEDs" on page
ing design" on page
Fig. 19: Plug terminal X4, potentiometric transducer pin assignment; TEDS
optional, view from below (solder side)
Important: Connect the sense leads to the corresponding excitation
voltage leads by wire bridges (2 at 2' and 3 at 3', also referred to as
jumpers or feedback bridges) if you are not using a 6-wire configura-
tion (with 5 wires assigned). Otherwise a sensor error will be sig-
naled, and you will not be able to measure (invalid measured value).
46
23.
Pt100
2
TEDS
1
S
Inner cable shield
Measurement signal -
4
1
Measurement signal +
Sense lead -
2'
2
Excitation voltage -
3'
Sense lead +
3
Excitation voltage +
Outer cable shield
AI
I
U
A04643_04_X00_03 HBM: public
24,
1-Wire-TEDS
1
"Shielding and ground-
2
3
ClipX

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading