Measuring channel objects

General channel settings
As subindex 0 is not usable as a parameter, the indices here extend
from 1 to 32 and not from 0 to 31 as in the signal references table. So
you always have to add 1 to the values in the
ences" on page
nal for example.
Subin-
Index
dex
0x44F0
3 ... 32
See
"List
of signal
0x44F1
refer-
ences" on
page 222
0x44F2
+ 1
0x44F3
0x44F4
1
184
222. Enter 4 for the gross signal and 5 for the net sig-
Function
Read (measurement) val-
ues.
Read or write channel
name.
Read or write unit.
Number of one decimal
places; only used for dis-
play in web browser.
Measurement Status of all
32 (measurement) values:
Bit = 0: OK
Bit = 1: Value is invalid
2 ... 31, bit position of the
value in question – see
of signal references" on
page
222.
A04643_04_X00_03 HBM: public
"List of signal refer-
Value
FLOAT, RO.
STRING, length ≤ 22 charac-
ters.
STRING, length ≤ 10 charac-
ters, parameters 1 ... 10 and 32
RO.
0 ... 6, UINT8.
32
0 ... 2
-1, UINT32, RO.