Example: Sine with amplitude 50 and offset 50
The amplitude fluctuates sinusoidally between 0 and 100 at the set frequency.
If the result is outside the number range, NaN (not a number) is outputted.

Counter function

At an offset of 0 the counter generates a number between -Amplitude and
+Amplitude that is incremented by one each time by the update rate (1000 Hz).
Any entered frequency is ignored.
White noise function
The noise signal is generated by a hardware noise source (integrated circuit).
7.4.2.14 Pulse-width measurement
This calculation measures the time between two edges. You can use one or two
(digital) channels (flags) as the input. The result can be output as time (seconds
or milliseconds) or as frequency (1/s). Specify the same channel for Start with
and Stop with, and use different levels if you only want to use one input chan-
nel.
Maximum resolution (smallest measurement duration): 1/update rate; at an
update rate of 1000/s this is equivalent to 1 ms.
Maximum measurement duration: 600 s. The output becomes invalid and the
calculation is stopped if the maximum measurement duration is exceeded. This
status is reset by the next start signal.
Pulse duration, corresponding frequency and achievable measurement uncer-
tainty
Pulse duration in ms
10
20
50
ClipX
Frequency in Hz
100
50
20
A04643_04_X00_03 HBM: public
Starting up the ClipX
Measurement uncer-
tainty as a %
10
5
2
101