Starting up the ClipX
Low flags. The values for Maximum Value, Minimum Value, Peak-to-Peak
and Mean are additionally available.
1
0
1
0
Duration
Fig. 39: Tolerance window; Monitoring for underflow and overflow
See also
"Peak with capture"
7.4.2.3
Peak with capture
This calculation determines the minimum, maximum or oscillation bandwidth of
a signal. You can also have the value of another channel (Hold Channel) deter-
mined when an extreme value is reached (output signal Captured Value). You
can tell by the falling edge of the peak flag (about 1 ms after detection of a new
peak value) that a new peak value has been found. As long as the measured
value continuously rises (in case of maximum) or falls (in case of minimum) the
peak flag is High (1).
Select the desired function and if necessary create a digital input to reset the
value (level-controlled, in case of High (1), Reset by). Set Reset on to Low
level to reset as long as a Low signal (0) is active.
90
Threshold High = 1 → NOK
Input OK
Threshold High
Maximum Value
Minimum Value
Threshold Low
Start
Stop
t
below.
A04643_04_X00_03 HBM: public
Peak-to-Peak