Herunterladen Inhalt Inhalt Diese Seite drucken

HBM ClipX Bedienungsanleitung Seite 62

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Electrical connections, LEDs
Important: On the first device you must enable Sync Mode Master
[Master sync mode]. All the others operate as slaves, so the switch
must not be active. Make the setting in your browser using the Ampli-
fier and Sensor Type menu items. Synchronization is only possible
for sensor types with carrier frequency (CF), not for DC. No synchro-
nization of the time base or the A/D converters takes place.
CxA
CxB
Sync
DI2
DI1
DO2
DO1
0V
24V
Fig. 30: Plug terminal X2, pin assignment for Sync signal
Important: Termination resistors are not necessary, and must not be
used. The maximum cable length between two devices is 30 cm.
60
Housing, cable shields
ClipX bus
X
ClipX bus GND
ClipX bus
Synchronization
DI
Digital In GND
Digital In 2
Digital In 1
Digital Out 2
Digital Out 1
Supply 0 V
Supply 10 ... 30 V
A04643_04_X00_03 HBM: public
CxA
X
CxB
Sync
DI
DI2
DI1
DO2
DO1
0V
24V