Teds Status: List Of Status Bits - HBM ClipX Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Status/error
Test signal active
Ethernet connection
established
Reserved
Reading TEDS
Error TEDS
Heartbeat
9.6.3

TEDS status: List of status bits

The following table sets out the bits in the TEDS status (32 bits) if the status in
question is set.
Status
No TEDS found
1-Wire
Zero-Wire
Invalid data
TEDS active/OK
ClipX
Bit
Explanation for bit set
20
The test signal is activated, no measured values are
captured.
21
An Ethernet connection has been established. This
is not the connection to a browser, it is the connec-
tion to a computer or PLC, e.g. for transferring data
or commands.
22 ...
Not in use, and reserved for future upgrades.
28
29
The TEDS module is being read and the device is
being configured according to the specified settings.
30
The data in the TEDS module either contain errors
or cannot be set.
31
The bit is switched on and off rhythmically at 0.5 Hz,
indicating that the ClipX is working.
Bit
0
Check that the TEDS module is correctly connected,
if one is installed.
See also
2
The ClipX has detected a 1-Wire TEDS module.
3
The ClipX has detected a Zero-Wire TEDS module.
4
The TEDS module contains invalid data. Check the
entries in the TEDS module (requires different hard-
ware).
5
All supported templates in the TEDS module have
been read and the parameters are set.
A04643_04_X00_03 HBM: public
Explanation
"Connecting transducers" on page
Operation via fieldbus
36.
161