HBM ClipX Bedienungsanleitung Seite 155

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

You can also choose more than one option. If you do, the settings are activated
in the following order:
1. The ClipX is trying to obtain an address from the DHCP server.
2. The ClipX is trying to obtain an address via BOOTP.
3. The ClipX activates the static IP address.
You must choose at least one of the options, otherwise you will get an error
message. Click on
Check the status either via the LEDs
in the web browser by the indicator next to Status at the top right in the window:
Modbus: NO CONFIG
Modbus: IDLE
Modbus: RUN
Register for transferring from ClipX to Modbus (input)
See also
Dis-
Input
crete
register
input
0 ... 31
0 ... 1
32 ... 95
2 ... 5
ClipX
APPLY
to activate the setting(s).
"Data transfer from the ClipX to the controller" on page
Meaning
Return value of the control word.
See
"The control word" on page
Coil 0 contains bit 0 'Zero'.
IO flags.
See
"Digital flags: List of I/O flags (I/O status)"
on page
162.
Coil 32 contains flag 0 'Digital input 1'.
A04643_04_X00_03 HBM: public
("Health-Monitoring, LEDs" on page
The ClipX is waiting for an IP address.
The ClipX is waiting for a connection.
The ClipX is connected.
Operation via fieldbus
Format
Bit array
163.
Bit array
26) or
129.
153