Operation via fieldbus
TEDS
Subindices 74 to 76 are only available with firmware version 2.02 or
higher.
Subin-
Index
dex
0x4300
2
3
4
11
12
13
14
15
16
17
18
190
Function
Parameters for usage type.
Status details
Bits see
"TEDS status: List
of status bits" on page
Command "Search and Use
TEDS as appropriate"; the
command is dependent on
the usage type.
Manufacturer.
Manufacturer.
Model number.
Model.
Measurement Location ID.
Serial number.
Version number.
Version Letter.
A04643_04_X00_03 HBM: public
Value
0: Ignore TEDS
1: Use TEDS, if available
2: TEDS required, UINT8.
32
0 ... 2
-1, UINT32, RO.
161.
WO.
Manufacturer ID, UINT32, RO.
Name of the manufacturer,
STRING, RO.
Sensor type ID, UINT32, RO.
Sensor type/model as text,
STRING, RO.
Measurement location ID,
UINT32, RO.
Serial number, UINT32, RO.
Version number, UINT32, RO.
Version letter, STRING, RO.