Electrical connections, LEDs
own device is to appear) in your browser (ClipX Bus menu). Address 1 serves
as the bus master; all the other addresses are slaves. Specify which signal to
transmit at each device. The status (valid/invalid) is transmitted along with the
measurement signal. 1000 values per second are transmitted per device
(including CRC check).
If you set Own Address for a device to 0, the ClipX bus will be deac-
tivated for that device, meaning no other devices are visible, and no
own signal can be transmitted.
CxA
CxB
Sync
DI2
DI1
DO2
DO1
0V
24V
Fig. 28: Plug terminal X2, pin assignment for ClipX bus
Important: Termination resistors are not necessary, and must not be
used. The maximum cable length between two devices is 30 cm.
58
Housing, cable shields
ClipX bus
X
ClipX bus GND
ClipX bus
Synchronization
DI
Digital In GND
Digital In 2
Digital In 1
Digital Out 2
Digital Out 1
Supply 0 V
Supply 10 ... 30 V
A04643_04_X00_03 HBM: public
CxA
X
CxB
Sync
DI
DI2
DI1
DO2
DO1
0V
24V