12

Diagnostics and error rectification, FAQs

12.1 Possible errors

For the ClipX to be ready, the LEDs must light in the relevant colors – see
"Health-Monitoring, LEDs" on page
device has (has been assigned) an IP address, whether the device itself was ini-
tialized without error, or whether the fieldbus of the BM40IE or BM40PB is work-
ing correctly.
Quick check
No LED on: Check power supply.
System LED does not turn green: check system LED states.
Fieldbus LED red: check fieldbus LED states.
Is the correct fieldbus set for the BM40IE?
Check the fieldbus configuration software on the master to ensure that all
addresses and parameters are correctly set.
If the values listed in the table are exceeded, an error is signaled, and the val-
ues are invalid.
Description
Analog input ±10 Volts
Analog input ±20 mA
Analog input 4 ...
20 mA
Measuring range of half
bridge DC 2.5 mV/V
Measuring range of half
bridge DC 5 mV/V
ClipX
Diagnostics and error rectification, FAQs
26. You can identify by this whether the
Minimum
Maximum
value
value
-10
-20
4
-2.5
-5
A04643_04_X00_03 HBM: public
Error
below
10
-12
20
-24
20
2
2.5
-3
5
-6
Error
Unit
above
12
V
24
mA
24
mA
3
mV/V
6
mV/V
229