Starting up the ClipX
define a range in which the signal is averaged and used as the zero value for
further measurements.
You have two options in each case for starting measurement and zeroing:
1. By a limit value.
2. By a digital signal.
You can also combine the methods, i.e. start measuring a mean value, min/max
etc. by way of a limit value and start zero measurement by a digital signal.
If you do not want monitor the start and stop of the calculation via a digital sig-
nal, set Enable by to 1 (internal constant) and Start on to High level for exam-
ple. The calculation is then performed whenever the conditions (Threshold or
Start Measure/Start Zeroing with) are met.
Maximum Value, Minimum Value, Peak-to-Peak and Mean are the values
determined during the measuring time via the signal at the input. The Offset is
the mean value over the Zero Tracking Time. The Status contains an identifier
for the current status of the calculation – see diagram and table below. The
Measured Value is continually updated throughout the time (current measure-
ment value).
After a restart (Device menu), a zero measurement is first performed with the
duration of Zero Tracking Time.