Herunterladen Inhalt Inhalt Diese Seite drucken

Use Teds - HBM ClipX Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.7

Use TEDS

When the ClipX is powered up, or a sensor is connected, the information in the TEDS
module is read and the ClipX is set accordingly (sensor type, scaling, bridge excitation
voltage, etc.).
You can use transducers with both 1-Wire and Zero-Wire TEDS (TEDS module in the
sense lines). If both TEDS connection variants are installed in the transducer, only the 1-
Wire module is used.
The TEDS symbol displayed in the browser does not mean that only the
information from the TEDS module is used. It indicates that a TEDS module
is present in the sensor. However, the settings made when connecting the
sensor or switching on the ClipX based on the information in the TEDS can
be overwritten afterwards via the browser or the field buses.
Specifying usage
You can choose between three TEDS usage variants via your browser's TEDS menu:
1.
Ignore TEDS.
2.
Use TEDS, if installed.
3.
TEDS required.
In the third variant, an error is signaled if no TEDS is found, or cannot be set as speci-
fied.
Supported templates (Information)
Strain gage (template: Strain Gage)
Full or half bridge sensor (template: Bridge Sensor)
Voltage (setting Volt, template: High Level Voltage Output Sensor)
Sensor with current output (settings Current, template: Current Loop Output Sen-
sor)
Potentiometer (template: Potentiometric Voltage Divider)
Pt100 (template: Resistive Temperature Detector)
Two-point scaling (always included in the respective sensor template)
Polynomial scaling (template: Calibration Curve)
HBM Channel Name (max. 45 ASCII characters)
ClipX
Electrical connections, LEDs
49