7.3

User Management

After connecting via the browser, you are initially at the Operator user level (Symbol
Click on the icon at the top right of the window to select a different user level: Mainte-
nance (Symbol
User management enables you to restrict access to the ClipX menus through a browser.
The default setting is for no password to be set and you can switch directly to a different
user level. But you can set a different password for both the Maintenance and Adminis-
trator user levels, thus restricting access to the settings.
At the Operator and Maintenance user levels, only the Home and Visualization menus
are accessible by default. At the Administrator user level all setting menus are accessi-
ble.
If no input is made at the Maintenance or Administrator user level, the user
level is reset to Operator after 30 minutes. The user level is also reset when
you close your browser and reconnect.
Forgotten your password?
You can temporarily disable user management by pressing and holding
down the Reset button on the front of the ClipX while logging in or changing
a password (any password is accepted, including a blank password field).
You will normally need a second colleague to press the button while you are
logging in.
See also
Setting passwords and rights
Log in to the Administrator or Maintenance user level and use Change Password: Select
User Level to choose the user level for which you want to set the password. The length
of the password is not actually limited, but it might be that only a specific number of
entered characters can be displayed depending on the window size. All characters are
allowed, including special characters. For security reasons, you must enter the pass-
word twice before you can activate it with
Specify the menu items to be accessible at the Administrator user level at the Mainte-
nance level by checking the relevant checkboxes. At the Operator user level no changes
can be made; only the Home and Visualization menu items can be displayed.