After connecting via the browser, you are initially at the Operator user level (Symbol
Click on the icon at the top right of the window to select a different user level: Mainte-
nance (Symbol
user levels, only the Home and Visualization menus are accessible by default. At the
Administrator user level all setting menus are accessible.
See also
„User Management" on page
After making the ClipX device settings in your browser, save them all to your PC
by choosing Device Storage from the menu.
7.4.1

Assistant to measure sensors

The Assistant menu option will help you to measure your sensor if you do not have a
calibration protocol or you want to measure the sensor when installed. Before measur-
ing, set the sensor type in the Amplifier menu to enable the ClipX to perform measure-
ments. Also specify a suitable low-pass filter, e.g. 10 Hz for static measuring, so as to
obtain the most stable possible measurement values. For dynamic measuring, you
must select the filter matching your signal frequencies and process.
The assistant provides two measurement variants:
1.
Mode: Static
Two points are measured: the value with the sensor under no load and the value
with the sensor under load (force/pressure/torque, etc.). The measurement itself is
performed over a period of about 6 seconds with the filter set in Amplifier. The cal-
culated mean value and the (sample) standard deviation are displayed both during
and after the measurement process.
2.
Mode: Dynamic
This mode is suitable when you are not able to apply a constant load (force/pres-
sure/torque, etc.). The peak values are calculated over your measuring time and
displayed.
You can either enter the physical value of the load manually or import it over the ClipX
bus or via one of the interfaces.
Procedure
1.
Select the Mode.
2.
Choose whether you want to enter the reference values manually or import them.