ERR
Status
LED
Off
off
Flashing at
2.5 Hz
Single flash
Red
Double flash
PROFINET LEDs (SF, BF, only BM40IE)
SF LED
Status
Off
off
Flashing
(1 Hz, 3 sec-
onds)
red
On
30
Meaning
No error.
The EtherCAT communication is running and error-free.
Invalid configuration.
Possible cause: A change specified by the master is not
possible.
Local error: The ClipX has autonomously changed the Eth-
erCAT status.
Possible causes:
– A host watchdog timeout has occurred,
– synchronization error. In this case the device switches
automatically to the SAFE-OPERATIONAL state.
A process data watchdog timeout has occurred.
Possible cause: A synchronization timeout (Sync Manager
watchdog).
Meaning (system error LED)
No error.
A DCP signal service is triggered over the bus.
Watchdog timeout.
There is a system error, or a channel, generic or extended
diagnosis.
Electrical connections, LEDs
ClipX