Either enter the reference values (the unit will be applied from the Amplifier menu,
though you can change it here) or specify the signal source for the reference values
(e.g. from the ClipX bus).
4.
Start the measurement.
In static measuring mode, the sensor is first measured under no load. Then you
must apply a load to the sensor and start the second measurement.
You can stop dynamic measuring mode when the minimum and maximum values
have been reached at least once.
5.
When the measurements are complete, you can either accept the measured values
APPLY
and click
The already measured values will then be set off against the new values (averag-
ing).
The window closes when you click
7.4.2

Calculation functions

Specify your calculations by choosing Calculated Channels from the menu. You can
create up to six calculations, and output up to six result channels plus eight digital flags
(the 6x6-matrix calculation already delivers up to six result channels). Each calculation
is executed 1000 times per second.
Procedure
1.
Click on one of the six lines at the top of the window under Function Blocks.
2.
Select one of the calculation functions under Select Function Type.
3.
Select the inputs or numerical values you want.
If you need more constants than those already available (internal constants), you
must first specify the number at the bottom of the window in the User-defined Con-
stants area. You can create up to 10 custom constants and assign them suitable
names.
The formula for calculating the respective function indicates how the variables x
x
etc. will be used. Check that all variables have corresponding values, even if they
2
are not used. Otherwise, a multiplication by 0 for example might make the preced-
ing variable practically ineffective.
4.
Select the channel in which you want the result(s) to be outputted. For digital signal,
eight flags (bits) are provided.
5.
Give the result channels informative names. Flags cannot be renamed.
80
or start a new measurement cycle.
APPLY
.
,
1