TEDS transducer identification

9
TEDS transducer identification
24
TEDS (Transducer Electronic Data Sheet) allows you to
store the sensor characteristic values in a chip as per the
IEEE 1451.4 standard. The C2 can be delivered with
TEDS, which is then fitted in the transducer housing,
connected and supplied with data by HBM before deliv­
ery. In this case, the results of calibration by HBM are
documented in the manufacturing certificate and saved
electronically in TEDS.
The TEDS module is connected between PIN E (sense
lead (-)) and PIN D (excitation voltage lead (-)). HBM's
zero wire configuration allows the TEDS to be read with­
out an additional sense lead.
If a suitable amplifier is connected (e.g. PMX from HBM),
then the amplifier electronics will read the TEDS chip and
parameterization will follow automatically, without any
intervention required by the user.
The chip content can be edited and modified with suitable
hardware and software. This can be implemented, e.g.
with the Quantum Assistant or even the DAQ software
CATMAN from HBM. Please comply with the operating
instructions of these products.