Herunterladen Inhalt Inhalt Diese Seite drucken

HBM ClipX Bedienungsanleitung Seite 43

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Strain gage full or half bridge and piezo-resistive sensors in 4-wire configuration with
Zero-Wire TEDS, version for D-sub-HD15
Important: The solder terminals on this TEDS module are not labeled. On the
side with the four contact surfaces (in the diagram above) you must connect
the adjacent contacts (short circuit or feedback bridges), otherwise the
measured value is always invalid. The contacts represented with dashed
lines are on the bottom of the TEDS module. Connect the + and - sense leads
there.
Pt100
TEDS
4
1
2'
2
3'
3
I
U
Abb. 21: Plug terminal X4, pin assignment in 4-wire configuration with Zero-Wire TEDS,
version for D-sub-HD15; with half bridge, connector 4 is omitted
ClipX
Electrical connections, LEDs
S
Inner cable shield
Measurement signal −
Measurement signal +
Sense lead −
Bridge excitation voltage −
Sense lead +
Bridge excitation voltage +
Outer cable shield
AI
Zero-Wire-TEDS
Cable wire colors (HBM transducer):
bu = blue; rd = red; bk = black; wh = white;
rd
wh
2
bk
1
bu
3
4
41