Herunterladen Inhalt Inhalt Diese Seite drucken

HBM ClipX Bedienungsanleitung Seite 46

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.6.4
Potentiometric transducer
See also
„Available connections and LEDs" on page
design" on page
22.
Abb. 24: Plug terminal X4, potentiometric transducer pin assignment; TEDS optional,
view from below (solder side)
Important: Connect the sense leads to the corresponding excitation voltage
leads by wire bridges (2 at 2' and 3 at 3', also referred to as jumpers or feed-
back bridges) if you are not using a 6-wire configuration (with 5 wires
assigned). Otherwise a sensor error will be signaled, and you will not be able
to measure (invalid measured value).
44
Pt100
2
TEDS
1
S
Inner cable shield
Measurement signal -
4
1
Measurement signal +
Sense lead -
2'
2
Excitation voltage -
3'
Sense lead +
3
Excitation voltage +
Outer cable shield
AI
I
U
24,
„Shielding and grounding
1-Wire-TEDS
1
2
3
Electrical connections, LEDs
ClipX