Status/error
Error parameter set
Error file system
Error ADC communi-
cation
Error ADC IRQ
Error ADC frozen (no
change for more
than 50 ms)
Error ADC DMA
Error DAC communi-
cation
DAC alarm
Error 1-wire com-
munication
Error ClipX bus
Error external RAM
Error sensor exci-
tation
Fieldbus I/O
Error fieldbus con-
troller
154
Bit
Explanation for bit set
5
Error in the current loaded parameter set. Load a
different parameter set or check all settings and
resave the parameter set. If the parameter set is
stored on PC, you can also import it from there and
check the stored version for errors.
6
Internal error in the device. Make a note of the error
type and contact HBM
7
231.
8
9
10
11
12
No current can flow at the current output; there is a
line break.
13
The 1-wire TEDS cannot be read. Check the wiring.
If possible, check whether the TEDS module can be
read on another device, or is defective.
14
The ClipX bus is not working correctly. Check the
wiring of the bus system.
15
Error in the RAM of the ClipX (not in the RAM of the
CPU). Make a note of the error type and contact
HBM
„Technical support" on page
16
The excitation voltage for the sensor has been
short-circuited. Check the wiring of the sensor.
17
Cyclic communication is taking place on the field-
bus (only BM40IE and BM40PB).
18
Internal error in the fieldbus controller (only on
BM40IE and BM40PB). Make a note of the error type
and contact HBM
„Technical support" on page
231.
„Technical support" on page
231.
ClipX
Operation via fieldbus