Herunterladen Inhalt Inhalt Diese Seite drucken

Diagnostics And Error Rectification, Faqs; Possible Errors - HBM ClipX Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ClipX:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

D
12
IAGNOSTICS AND ERROR RECTIFICATION
12.1

Possible errors

For the ClipX to be ready, the LEDs must light in the relevant colors – see
toring, LEDs" on page
assigned) an IP address, whether the device itself was initialized without error, or
whether the fieldbus of the BM40IE or BM40PB is working correctly.
Quick check
No LED on: Check power supply.
System LED does not turn green: check system LED states.
Fieldbus LED red: check fieldbus LED states.
Is the correct fieldbus set for the BM40IE?
Check the fieldbus configuration software on the master to ensure that all
addresses and parameters are correctly set.
If the values listed in the table are exceeded, an error is signaled, and the values are
invalid.
Description
Analog input ±10 Volts
Analog input ±20 mA
Analog input 4 ... 20 mA
Measuring range of half
bridge DC 2.5 mV/V
Measuring range of half
bridge DC 5 mV/V
Measuring range of half
bridge CF 2.5 mV/V
Measuring range of half
bridge CF 5 mV/V
ClipX

Diagnostics and error rectification, FAQs

26. You can identify by this whether the device has (has been
Minimum
value
-10
-20
4
-2.5
-5
-2.5
-5
, FAQ
Maximum
Error
value
below
10
-12
20
-24
20
2
2.5
-3
5
-6
2.5
-3
5
-6
S
„Health-Moni-
Error
Unit
above
12
V
24
mA
24
mA
3
mV/V
6
mV/V
3
mV/V
6
mV/V
225

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis