Herunterladen Inhalt Inhalt Diese Seite drucken

Mechanical Connection Variants - ECOLAB LMIT09 DPV1 Betriebsanleitung

Leitfähigkeits- und temperaturmessung
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Installation
6.1

Mechanical connection variants

CAUTION!
All mechanical installation work must only be performed by authorised,
trained specialists in accordance with local regulations. All connection
elements must be checked and, if necessary, retightened (it is possible for
them to become loose due to major fluctuations in the ambient temperature
or as a result of vibrations).
A
E
Fig. 2: MAN034247_6-0
A
Compact version, pipe installation, clamping ring
attachment
B
Wall-mounted unit
C
Compact version, pipe installation, flange connection
D
Type: Flow fitting (DIN 11850)
Material: 1,4404, EPDM seal
E
Type: Weld-on-ring for tank wall mounting
Material: 1,4404, EPDM seal
Transmitter with integrated transducer (compact version)
CAUTION!
The measuring bore in the measuring calotte must be aligned parallel to
the pipe's axis, i.e. in the flow direction of the medium. Deviations can
lead to distorted readings (eddy formation). The installation position can be
checked using the marking notches on the measuring calotte.
Installing the clamping ring version
Insert the transmitter with integrated transducer into the flow fitting or welded-on ring and
tighten using the clamping rings.
Installing the flange version
Insert the transmitter with integrated transducer into the flow fitting or welded-on flange in
the case of tank wall installation and secure with 4 screws.
417101561 Rev. 4-06.2021
B
F
34
C
G
F
Type: Weld-on-ring for pipe mounting
Material: 1,4404, EPDM seal
G
Type: Weld-on-flange for tank wall mounting
Material: 1,4404, EPDM seal
H
Type: Flow fitting
Material: 1,4404, EPDM seal
D
H

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Lmit09S209

Inhaltsverzeichnis