ECOLAB LMIT09 DPV1 Betriebsanleitung Seite 167

Leitfähigkeits- und temperaturmessung
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation
Current Output Cond. (0 or 4 mA...20 mA, measured value stability & alarm
level)
C
4
Fig. 18: Basic Settings - Current Output Cond.
1.
Call the main menu (item A) and open Basic Settings (item B)
Ä 8.4 'Main menu' on page 46).
(see
ð The Basic Settings menu is displayed.
2.
Press the 'Enter'
ð This selects the Language submenu.
3.
Press the 'Down'
ð This selects the Current Output Cond. submenu (item C).
4.
Press the 'Enter'
ð The Current Output Cond. submenu opens.
The display shows the Range (item 4) for configuration.
5.
Use the buttons:
Setting range: 0 or 4 mA (factory setting: 4 mA)
6.
Press the 'Right'
Setting range: OFF or 1–10 seconds (factory setting: OFF).
7.
Use the buttons:
8.
Press the 'Right'
9.
Use the buttons:
Setting range: 0 mA / 2.4 mA or 22 mA (factory setting: 0 mA).
10. Press the 'Right'
417101561 Rev. 4-06.2021
B
A
1
2
5
button (item 1).
arrow button twice (item 3).
button (item 3).
or
to set the Range (item 4).
arrow button to set the Stability (item 5).
NOTICE!
When the 'Stability' setting is selected, an average measured value is
calculated over the indicated time. It is used to stabilise the display
and the current output if the measurement is fluctuating. OFF = no
stability.
or
to set the Stability.
arrow button to access the Alarm Level setting (item 6).
or
to set the Alarm Level.
NOTICE!
If 0 mA – 20 mA is selected as the range then it is not permissible to
select 2.4 mA (in this case, the background changes to red after the
button is pressed).
arrow button to access the Monitoring setting (item 7).
6
54
3
7
8