ECOLAB LMIT09 DPV1 Betriebsanleitung Seite 194

Leitfähigkeits- und temperaturmessung
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1
Fig. 44: Calibration - Zero Point Correction
1.
Select the menu (
ð Calibration is called and Armature Factor is shown with a grey
background (inverted).
2.
Press the 'Down'
ð The Zero Correction submenu is displayed in grey (inverted).
3.
Press the 'Enter'
ð The zero point measurement is displayed and automatically performed in all five
measuring ranges.
4.
After successful zero point calibration, the display switches from BUSY to OK.
5.
Press the
button (item 5) to return to the Calibration menu.
You can use the
Ä 'Select Calibration' on page 78).
arrow button (item 1) twice.
button (item 2) to start the zero point measurement.
button (item 4) to repeat the Calibration.
81
2
5
417101561 Rev. 4-06.2021