ECOLAB LMIT09 DPV1 Betriebsanleitung Seite 170

Leitfähigkeits- und temperaturmessung
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

NOTICE!
With current output monitoring ON, the device monitors the
conductivity/concentration output current range from 1 mA to 20 mA.
When a residual current occurs in the conductivity/concentration loop,
an alarm is shown on the display and the alarm relay is activated (e.g.
in the case of a line break).
Residual currents that flow outside the LMIT09 DPV1 (e.g. via PE
loop) cannot be monitored. OFF = no monitoring.
12. You can use the
ð All settings are saved and the display returns to Basic settings menu (see
Ä 8.5 'Basic settings' on page 49).
CIP Functions OFF, Signal A (CIP 1/2)) only, Signal A&B (CIP 1- 4)
This menu item is used to activate the sets of parameters (CIP ranges).
Every set of parameters (CIP range) can be individually adjusted.
A
C
4
Fig. 20: Basic Settings – CIP Functions
1.
Call the main menu (item A) and open Basic Settings (item B)
Ä 8.4 'Main menu' on page 46).
(see
ð The Basic Settings menu is displayed.
2.
Press the 'Enter'
ð This selects the Language submenu.
3.
Press the 'Down'
ð This selects the CIP Functions submenu (item C).
4.
Press the 'Enter'
ð The CIP Functions submenu opens.
The display shows the OFF selection (item 4) for configuration.
Select Off to deactivate the CIP Functions.
5.
Press
to accept the setting and exit the setting level.
ð All settings are saved and the display returns to the CIP Functions menu.
6.
To select Signal A (item 5) or Signal A&B (item 7), use the 'Up'
buttons.
Use the Signal A area to set the Range (item 6) as CIP 1 & 2.
Use the Signal A&B area to set the Range (item 8) as CIP 1–4.
7.
You can use the
ð All settings are saved and the display returns to Basic settings menu (see
Ä 8.5 'Basic settings' on page 49).
button (item 8) to confirm the set values at any time.
B
1
2
5
6
button (item 1).
arrow button twice (item 5).
button (item 3).
button (item 8) to confirm the set values at any time.
57
Operation
3
7
8
9
or 'Down'
417101561 Rev. 4-06.2021