Customer Table
The option 'Customer Table' is used to delete all saved customer product data. If a
customer product table is saved in a set of parameters, this set of parameters is
automatically switched to conductivity measurement.
1
2
Fig. 52: Service – Factory defaults – Customer Table
1.
Select the menu (
ð Factory defaults is called and Only Settings (item 1) is displayed in grey
(inverted).
2.
Press the 'Down'
ð The Customer Table submenu (item 3) is displayed in grey (inverted).
3.
Press the 'Enter'
ð The screen displays a confirmation prompt, which is set to NO by default (item 5)
to prevent data from being deleted accidentally.
4.
Use the arrow buttons (item 6) to switch between No (item 5) and Yes (item 7).
5.
Use the
button (item 8) to confirm the required setting.
ð
Choosing Yes (item 5) deletes the data.
The screen returns to the factory defaults, as described under
Manual Functions - Factory defaults' on page 85.
3
Ä 8.8 'Service' on page 83).
arrow button (item 2) twice.
button (item 4) to call the Customer Table selection screen.
87
5
4
8
Ä 'Calling