ECOLAB LMIT09 DPV1 Betriebsanleitung Seite 168

Leitfähigkeits- und temperaturmessung
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

11. Use the buttons:
Setting range: OFF or ON (factory setting: OFF)
NOTICE!
With current output monitoring ON, the device monitors the
conductivity/concentration output current range from 1 mA to 20 mA.
When a residual current occurs in the conductivity/concentration loop,
an alarm is shown on the display and the alarm relay is activated (e.g.
in the case of a line break).
Residual currents that flow outside the LMIT09 DPV1 (e.g. via PE
loop) cannot be monitored. OFF = no monitoring.
12. You can use the
ð All settings are saved and the display returns to Basic settings menu (see
Ä 8.5 'Basic settings' on page 49).
or
to set the Monitoring.
button (item 8) to confirm the set values at any time.
55
Operation
417101561 Rev. 4-06.2021