Manual cleaning with ultrasound and immersion disinfection

Phase
Step
I
Ultrasonic cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*Recommended: BBraun Stabimed fresh
Note the information on appropriate cleaning brushes and disposable
syringes, see Validated cleaning and disinfection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for
at least 15 min. Ensure that all accessible surfaces are immersed and
acoustic shadows are avoided.
Clean the product with a suitable cleaning brush in the solution until
all discernible residues have been removed from the surface.
If applicable, brush through non-visible surfaces with an appropriate
cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during
cleaning.
Thoroughly rinse through these components with the cleaning disin-
fectant solution (at least five times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under run-
ning water.
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
Drain any remaining water fully.
Phase III
Fully immerse the product in the disinfectant solution.
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
Rinse lumens at least five times at the beginning of the exposure time
with an appropriate disposable syringe. Ensure that all accessible sur-
faces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces) under run-
ning water.
Mobilize non-rigid components, such as set screws, joints, etc. during
final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air), see Validated cleaning and disinfection procedure.
14
D
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
RT (cold)
5
2
RT (cold)
1
-
RT
-
-
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
D–W
-
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
FD–W
-
-
-