Manual Cleaning/Disinfection; Manual Cleaning With Immersion Disinfection - B. Braun Aesculap OP950 Gebrauchsanweisung/Technische Beschreibung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
Aesculap
LED light source OP950 with sterile adapter OP941
8.7

Manual cleaning/disinfection

Prior to manual disinfecting, allow water to drain off for a sufficient
length of time to prevent dilution of the disinfectant solution.
After manual cleaning/disinfection, check visible surfaces visually for
residues.
Repeat the cleaning/disinfection process if necessary.
8.7.1

Manual cleaning with immersion disinfection

Phase
Step
I
Disinfecting cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable
syringes, see Validated cleaning and disinfection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15
min. Ensure that all accessible surfaces are moistened.
Clean the product with a suitable cleaning brush in the solution until
all discernible residues have been removed from the surface.
If applicable, brush through non-visible surfaces with an appropriate
cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during
cleaning.
Thoroughly rinse through these components with the cleaning disin-
fectant solution (at least five times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under run-
ning water.
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
Drain any remaining water fully.
Phase III
Fully immerse the product in the disinfectant solution.
12
®
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) >15
2
RT (cold) 1
-
RT (cold) 15
2
RT (cold) 1
-
RT
-
-
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
D–W
-
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
FD-W
-
-
-
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
Rinse lumens at least 5 times at the beginning of the exposure time
using an appropriate disposable syringe. Ensure that all accessible sur-
faces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces).
Mobilize non-rigid components, such as set screws, joints, etc. during
final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air), see Validated cleaning and disinfection procedure.