en
3.8.2
Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning with immersion
disinfection
■
SC410R
■
SC420R-SC426R
■
SC428R
■
SC430R-SC434R
■
FW076R
Manual cleaning with ultrasound
and immersion disinfection
■
SC429R
■
SC436R
Mechanical alkaline cleaning and
thermal disinfection
■
SC410R
■
SC420R-SC421R
■
SC430R–SC434R
■
FW076R
Manual pre-cleaning with brush
and subsequent mechanical alka-
line cleaning and thermal disinfec-
tion
■
SC422R-SC426R
■
SC428R
Manual pre-cleaning with ultra-
sound and brush, and subsequent
mechanical alkaline cleaning and
thermal disinfection
■
SC429R
■
SC436R
12
Specific requirements
■
Suitable cleaning brush: e.g. TA011327 for SC422R and SC428R,
FO764200 for SC429R
■
20 ml disposable syringe
■
Keep working ends open for cleaning.
■
When cleaning instruments with movable hinges, ensure that
these are in an open position and, if applicable, move the hinge
while cleaning.
■
Drying phase: Use a lint-free cloth or medical compressed air
■
20 ml disposable syringe
■
Drying phase: Use a lint-free cloth or medical compressed air
■
Place the product on a tray that is suitable for cleaning (avoid
rinsing blind spots).
■
Connect components with lumens and channels directly to the
rinsing port of the injector carriage.
■
Keep working ends open for cleaning.
■
Place the product on the tray with all product links and joints
open.
■
Suitable cleaning brush: e.g. TA011327 for SC422R and SC428R
■
20 ml disposable syringe
■
Place the product on a tray that is suitable for cleaning (avoid
rinsing blind spots).
■
Connect components with lumens and channels directly to the
rinsing port of the injector carriage.
■
Keep working ends open for cleaning.
■
Place the product on the tray with all product links and joints
open.
■
Suitable cleaning brush: e.g. FO764200 for SC429R and SC436R
■
20 ml disposable syringe
■
Place the product on a tray that is suitable for cleaning (avoid
rinsing blind spots).
■
Connect components with lumens and channels directly to the
rinsing port of the injector carriage.
■
Keep working ends open for cleaning.
■
Place the product on the tray with all product links and joints
open.
Reference
Chapter Manual cleaning/disinfection and
subsection:
■
Chapter Manual cleaning with immer-
sion disinfection
Chapter Manual cleaning/disinfection and
subsection:
■
Chapter Manual cleaning with ultra-
sound and immersion disinfection
Chapter Mechanical cleaning/disinfection
and subsection:
■
Chapter Mechanical alkaline cleaning
and thermal disinfecting
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and subsection:
■
Chapter Manual pre-cleaning with a
brush
■
Chapter Mechanical alkaline cleaning
and thermal disinfecting
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and subsection:
■
Chapter Manual pre-cleaning with
ultrasound and brush
■
Chapter Mechanical alkaline cleaning
and thermal disinfecting