Validated cleaning and disinfection procedure

Validated procedure
Manual cleaning with immersion
disinfection
SC410R
SC420R-SC426R
SC428R
SC430R-SC434R
FW076R
Manual cleaning with ultrasound
and immersion disinfection
SC429R
SC436R
Mechanical alkaline cleaning and
thermal disinfection
SC410R
SC420R-SC421R
SC430R–SC434R
FW076R
Manual pre-cleaning with brush
and subsequent mechanical alka-
line cleaning and thermal disinfec-
tion
SC422R-SC426R
SC428R
Manual pre-cleaning with ultra-
sound and brush, and subsequent
mechanical alkaline cleaning and
thermal disinfection
SC429R
SC436R
12
Specific requirements
Suitable cleaning brush: e.g. TA011327 for SC422R and SC428R,
FO764200 for SC429R
20 ml disposable syringe
Keep working ends open for cleaning.
When cleaning instruments with movable hinges, ensure that
these are in an open position and, if applicable, move the hinge
while cleaning.
Drying phase: Use a lint-free cloth or medical compressed air
20 ml disposable syringe
Drying phase: Use a lint-free cloth or medical compressed air
Place the product on a tray that is suitable for cleaning (avoid
rinsing blind spots).
Connect components with lumens and channels directly to the
rinsing port of the injector carriage.
Keep working ends open for cleaning.
Place the product on the tray with all product links and joints
open.
Suitable cleaning brush: e.g. TA011327 for SC422R and SC428R
20 ml disposable syringe
Place the product on a tray that is suitable for cleaning (avoid
rinsing blind spots).
Connect components with lumens and channels directly to the
rinsing port of the injector carriage.
Keep working ends open for cleaning.
Place the product on the tray with all product links and joints
open.
Suitable cleaning brush: e.g. FO764200 for SC429R and SC436R
20 ml disposable syringe
Place the product on a tray that is suitable for cleaning (avoid
rinsing blind spots).
Connect components with lumens and channels directly to the
rinsing port of the injector carriage.
Keep working ends open for cleaning.
Place the product on the tray with all product links and joints
open.
Reference
Chapter Manual cleaning/disinfection and
subsection:
Chapter Manual cleaning with immer-
sion disinfection
Chapter Manual cleaning/disinfection and
subsection:
Chapter Manual cleaning with ultra-
sound and immersion disinfection
Chapter Mechanical cleaning/disinfection
and subsection:
Chapter Mechanical alkaline cleaning
and thermal disinfecting
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and subsection:
Chapter Manual pre-cleaning with a
brush
Chapter Mechanical alkaline cleaning
and thermal disinfecting
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and subsection:
Chapter Manual pre-cleaning with
ultrasound and brush
Chapter Mechanical alkaline cleaning
and thermal disinfecting