Herunterladen Inhalt Inhalt Diese Seite drucken

Mechanical Cleaning/Disinfection; Mechanical Alkaline Cleaning And Thermal Disinfecting - B. Braun Aesculap Spine Quintex SC420R Gebrauchsanweisung/Technische Beschreibung

Anteriores zervixplattensystem - instrumente
Vorschau ausblenden Andere Handbücher für Aesculap Spine Quintex SC420R:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22

3.10 Mechanical cleaning/disinfection

Note
The cleaning and disinfection device must be of tested and approved effec-
tiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced
and checked at regular intervals.

3.10.1 Mechanical alkaline cleaning and thermal disinfecting

Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
DW:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechanical cleaning/disinfect-
ing.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD-W
>10/50
1
FD–W
90/194
5
FD-W
-
-
-
Chemistry/Note
-
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH ~ 11*
-
-
According to the program for cleaning and disinfection device
en
15

Werbung

Inhaltsverzeichnis
loading