12. Minimum Requirements for Additional Components

12.1. Devices

Endoscope camera unit:
Outputs: DVI / SDI / HD-SDI / 3G-SDI / S-VIDEO / VGA
Resolution: minimum PAL 768 x 572 - maximum 1080p60
Endoscopic image delay in combination with 60 Hz medical monitor: maximum 50 ms
Camera unit according to IEC 6060-1 and provides 2 MOPP to the camera
Lightsource according to IEC 60601-1 and provides 2 MOPP towards the patient
Camera head:
Resolution: PAL (752(H) x 582(V)) or higher
Endoscopes:
Diameter: 4 mm
Visual angle: 0° / 30° / 45°
Default bayonet connector
Medical monitors:
Inputs: DVI or HDMI
Resolution: 1280 x 1024 or higher, 1920 x 1080 recommended
Colors: 16.7 million colors
Maximum delay of 50 ms in combination with an endoscope camera unit
Medical grade
Fulfills IEC 60950-1 or IEC 60601-1
The device is compatible with the following monitors:
240-031-020 VisionPro
240-031-050 4K Display