Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker ENT Bedienungsanleitung Seite 31

Navigation system
Vorschau ausblenden Andere Handbücher für ENT:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 75
Environmental
Limitations
Temperature
Relative Humidity
Atmospheric Air Pressure
Essential Performance Features:
The product detects and visualizes defined navigation positions with an accuracy of at
least 2 mm in radiological images.
The delay produced by the product in processing the incoming real-time video image
data must not exceed 250 ms.
No undetected data corruption.
No unnoticed freezing of the instrument display.
No unnoticed freezing of the incoming real-time video image data.
No faulty instrument assignment.
EN
Operation
Between 10 °C and 30 °C
30 % to 75 %
Between 80 kPa and 106 kPa
Storage and
Transportation
Between -10 °C and +50 °C
10 % to 90 %, non-condensing
Between 50 kPa and 106 kPa
Technical Specifications | 27/34