Environmental
Limitations
Temperature
Relative Humidity
Atmospheric Air Pressure
Essential Performance Features:
–
The product detects and visualizes defined navigation positions with an accuracy of at
least 2 mm in radiological images.
–
The delay produced by the product in processing the incoming real-time video image
data must not exceed 250 ms.
–
No undetected data corruption.
–
No unnoticed freezing of the instrument display.
–
No unnoticed freezing of the incoming real-time video image data.
–
No faulty instrument assignment.
EN
Operation
Between 10 °C and 30 °C
30 % to 75 %
Between 80 kPa and 106 kPa
Storage and
Transportation
Between -10 °C and +50 °C
10 % to 90 %, non-condensing
Between 50 kPa and 106 kPa
Technical Specifications | 27/34