Guidance and Manufacturer's Declaration - Electromagnetic Emission
The Electromagnetic Navigation Unit is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the product should ensure that it is used in
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC
6100–3–2
Voltage fluctuations /
flicker emissions IEC
6100–3–3
EN
such an environment.
Compliance
Group 1
Class A
Class A
Compliant
Electromagnetic Compatibility | 29/34
Electromagnetic Environment
–Guidance
The product uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause interference in nearby
electronic equipment.
The product is suitable for use
in other establishments as the
domestic establishments and
those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes, provided that the
following warning is fully con-
sidered:
WARNING
The product is solely intended
to be used by medical staff.
Within domestic conditions it
can cause radio interferences
which can demand taking cor-
rective action, for example,
new orientation, new arrange-
ment or shielding of the prod-
uct, or filtering of the connec-
tion to the location.